{"dataType":"CVE_RECORD","dataVersion":"5.1","cveMetadata":{"cveId":"CVE-2024-48973","assignerOrgId":"dba971b9-eb30-4121-91e1-3b45611354aa","state":"PUBLISHED","assignerShortName":"Baxter","dateReserved":"2024-10-10T19:24:41.495Z","datePublished":"2024-11-14T21:24:14.099Z","dateUpdated":"2024-11-18T15:33:44.924Z"},"containers":{"cna":{"affected":[{"defaultStatus":"unaffected","product":"Life2000 Ventilation System","vendor":"Baxter","versions":[{"status":"affected","version":"06.08.00.00 and prior"}]}],"datePublic":"2024-11-14T21:21:00.000Z","descriptions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.<br>"}],"value":"The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance."}],"impacts":[{"capecId":"CAPEC-117","descriptions":[{"lang":"en","value":"CAPEC-117 Interception"}]},{"capecId":"CAPEC-441","descriptions":[{"lang":"en","value":"CAPEC-441 Malicious Logic Insertion"}]}],"metrics":[{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"LOCAL","availabilityImpact":"HIGH","baseScore":9.3,"baseSeverity":"CRITICAL","confidentialityImpact":"HIGH","integrityImpact":"HIGH","privilegesRequired":"NONE","scope":"CHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-1263","description":"CWE-1263 Improper Physical Access Control","lang":"en","type":"CWE"}]}],"providerMetadata":{"orgId":"dba971b9-eb30-4121-91e1-3b45611354aa","shortName":"Baxter","dateUpdated":"2024-11-14T21:49:44.756Z"},"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"}],"source":{"discovery":"UNKNOWN"},"title":"Debug port on Life2000 Ventilator serial interface is enabled by default","x_generator":{"engine":"Vulnogram 0.2.0"}},"adp":[{"affected":[{"vendor":"baxter","product":"life2000_ventilator_firmware","cpes":["cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*"],"defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThanOrEqual":"06.08.00.00","versionType":"custom"}]}],"metrics":[{"other":{"type":"ssvc","content":{"timestamp":"2024-11-18T15:33:07.996857Z","id":"CVE-2024-48973","options":[{"Exploitation":"none"},{"Automatable":"no"},{"Technical Impact":"total"}],"role":"CISA Coordinator","version":"2.0.3"}}}],"title":"CISA ADP Vulnrichment","providerMetadata":{"orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","shortName":"CISA-ADP","dateUpdated":"2024-11-18T15:33:44.924Z"}}]}}