{"dataType":"CVE_RECORD","dataVersion":"5.1","cveMetadata":{"cveId":"CVE-2024-48974","assignerOrgId":"dba971b9-eb30-4121-91e1-3b45611354aa","state":"PUBLISHED","assignerShortName":"Baxter","dateReserved":"2024-10-10T19:24:41.495Z","datePublished":"2024-11-14T21:27:10.284Z","dateUpdated":"2024-11-18T15:32:38.488Z"},"containers":{"cna":{"affected":[{"defaultStatus":"unaffected","product":"Life2000 Ventilation System","vendor":"Baxter","versions":[{"status":"affected","version":"06.08.00.00 and prior"}]}],"datePublic":"2024-11-14T21:50:00.000Z","descriptions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure."}],"value":"The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure."}],"impacts":[{"capecId":"CAPEC-186","descriptions":[{"lang":"en","value":"CAPEC-186 Malicious Software Update"}]},{"capecId":"CAPEC-117","descriptions":[{"lang":"en","value":"CAPEC-117 Interception"}]}],"metrics":[{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"LOCAL","availabilityImpact":"HIGH","baseScore":9.3,"baseSeverity":"CRITICAL","confidentialityImpact":"HIGH","integrityImpact":"HIGH","privilegesRequired":"NONE","scope":"CHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-494","description":"CWE-494 Download of Code Without Integrity Check","lang":"en","type":"CWE"}]}],"providerMetadata":{"orgId":"dba971b9-eb30-4121-91e1-3b45611354aa","shortName":"Baxter","dateUpdated":"2024-11-14T21:50:36.915Z"},"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"}],"source":{"discovery":"UNKNOWN"},"title":"Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates","x_generator":{"engine":"Vulnogram 0.2.0"}},"adp":[{"affected":[{"vendor":"baxter","product":"life2000_ventilator_firmware","cpes":["cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*"],"defaultStatus":"unaffected","versions":[{"version":"0","status":"affected","lessThanOrEqual":"06.08.00.00","versionType":"custom"}]}],"metrics":[{"other":{"type":"ssvc","content":{"timestamp":"2024-11-18T15:31:47.610421Z","id":"CVE-2024-48974","options":[{"Exploitation":"none"},{"Automatable":"no"},{"Technical Impact":"total"}],"role":"CISA Coordinator","version":"2.0.3"}}}],"title":"CISA ADP Vulnrichment","providerMetadata":{"orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","shortName":"CISA-ADP","dateUpdated":"2024-11-18T15:32:38.488Z"}}]}}