{"dataType":"CVE_RECORD","dataVersion":"5.2","cveMetadata":{"cveId":"CVE-2025-14096","assignerOrgId":"46b595e9-1acc-41cb-9398-adaf98d37a9b","state":"PUBLISHED","assignerShortName":"Radiometer","dateReserved":"2025-12-05T10:50:00.566Z","datePublished":"2025-12-17T12:19:27.489Z","dateUpdated":"2025-12-17T14:43:16.523Z"},"containers":{"cna":{"affected":[{"defaultStatus":"unaffected","platforms":["Windows XP","Windows 7"],"product":"ABL90 FLEX and ABL90 FLEX PLUS Analyzers","vendor":"Radiometer Medical Aps","versions":[{"status":"affected","version":"Windows 7 Operating system","versionType":"Radiometer Customized Windows Operating System"},{"status":"affected","version":"Windows XP Operating system","versionType":"Radiometer Customized Windows Operating System"}]},{"defaultStatus":"unaffected","platforms":["Windows 7","Windows XP"],"product":"AQT90 FLEX Analyzers","vendor":"Radiometer Medical Aps","versions":[{"status":"affected","version":"Windows 7 Operating system","versionType":"Radiometer Customized Windows Operating System"},{"status":"affected","version":"Windows XP Operating system","versionType":"Radiometer Customized Windows Operating System"}]},{"defaultStatus":"unaffected","platforms":["Windows 7","Windows XP"],"product":"ABL800 BASIC and ABL800 FLEX analyzers","vendor":"Radiometer Medical Aps","versions":[{"status":"affected","version":"Windows 7 Operating system","versionType":"Radiometer Customized Windows Operating System"},{"status":"affected","version":"Windows XP Operating system","versionType":"Radiometer Customized Windows Operating System"}]}],"configurations":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"Requires physical access to the Analyzer.\n\n<br>"}],"value":"Requires physical access to the Analyzer."}],"credits":[{"lang":"en","type":"reporter","value":"Florian Hauser and Fabian Weber from CODE WHITE GmbH"}],"descriptions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.<br><br>Other related CVE's are CVE-2025-14095 &amp; CVE-2025-14097.<br><br>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.</span>\n\n<br><br>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Required Configuration for Exposure:<br></span>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Attacker requires physical access to the analyzer.</span>\n\n<br><br>Temporary work Around:<br>Only authorized people can physically access the analyzer.<br><br>Permanent solution:<br>Local Radiometer representatives will contact all affected customers to discuss a permanent solution.<br><br>Exploit Status:<br>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.</span>\n\n<br><br>"}],"value":"A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\n\nOther related CVE's are CVE-2025-14095 & CVE-2025-14097.\n\n\n\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\n\n\nRequired Configuration for Exposure:\n\n\nAttacker requires physical access to the analyzer.\n\n\n\nTemporary work Around:\nOnly authorized people can physically access the analyzer.\n\nPermanent solution:\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\nExploit Status:\n\n\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."}],"exploits":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\n\n<br>"}],"value":"Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."}],"impacts":[{"capecId":"CAPEC-69","descriptions":[{"lang":"en","value":"CAPEC-69 — Target Programs with Elevated Privileges"}]}],"metrics":[{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"LOCAL","availabilityImpact":"HIGH","baseScore":8.4,"baseSeverity":"HIGH","confidentialityImpact":"HIGH","integrityImpact":"HIGH","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-798","description":"CWE-798: Use of Hard-coded Credentials","lang":"en","type":"CWE"}]},{"descriptions":[{"cweId":"CWE-250","description":"CWE-250: Execution with Unnecessary Privileges","lang":"en","type":"CWE"}]}],"providerMetadata":{"orgId":"46b595e9-1acc-41cb-9398-adaf98d37a9b","shortName":"Radiometer","dateUpdated":"2025-12-17T13:03:05.696Z"},"references":[{"url":"https://www.radiometer.com/myradiometer"}],"solutions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"Local Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\n\n\n<br>"}],"value":"Local Radiometer representatives will contact all affected customers to discuss a permanent solution."}],"source":{"discovery":"EXTERNAL"},"timeline":[{"lang":"en","time":"2025-12-17T13:01:00.000Z","value":"CVE Published"}],"title":"Credential Disclosure vulnerability in Radiometer Products","workarounds":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"Only authorized people can physically access the analyzer.\n\n\n\n<br>"}],"value":"Only authorized people can physically access the analyzer."}],"x_generator":{"engine":"Vulnogram 0.5.0"}},"adp":[{"metrics":[{"other":{"type":"ssvc","content":{"timestamp":"2025-12-17T14:42:57.843796Z","id":"CVE-2025-14096","options":[{"Exploitation":"none"},{"Automatable":"no"},{"Technical Impact":"total"}],"role":"CISA Coordinator","version":"2.0.3"}}}],"title":"CISA ADP Vulnrichment","providerMetadata":{"orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","shortName":"CISA-ADP","dateUpdated":"2025-12-17T14:43:16.523Z"}}]}}